Specifications | Technique |
Capacity | X: 10/25/50 Y: 10/25/50 Z: 25/50/100 |
Sensitivity | ≈1.0 mV / V |
Zero Balance | ± 2% F.S. |
Non-linearity | 0.2% F.S. |
Hysteresis Error | 0.2% F.S. |
Repetition | 0.2% F.S. |
Interference error | 0.5% F.S. |
Creep (30 min) | 0.2% F.S. |
Temp. Effect on sensitivity (TKc) | 0.1% F.S. / 10 ℃ |
Temp. Effect on zero balance (TKo) | 0.1% F.S. / 10 ℃ |
Input Resistance | 700 ± 10 Ω |
Output Resistance | 700 ± 10 Ω |
Insulating Resistance | ≥5000 MΩ / 100VDC |
Excitation Voltage | 5 ~ 10 V |
Max. Excitation Voltage | 15 V |
Compensation Range | -10 ~ 40 ℃ |
Service Temperature Range | -20 ~ 75 ℃ |
Limit load (E L) | 150% F.S. |
Breaking load (E d) | 300% F.S. |
Cable Size | φ6x3000 mm |
Weight (G), approx. | 1.6 Kg |
IP Class | IP66 |